CLASSIC Series | TCM 0650 Coriolis Mass Flow Meter
The TRICOR CLASSIC Series TCM 0650 Coriolis Mass Flow Meter has a maximum mass flow rate of 650 kg/hr (24 lb/min) or 98 bbl/day and nominal size of DN4 (1/8”), and is rated for process pressures up to 200 bar (2900 psi). Standard accuracy for the TCM 0650 is 0.3%, but with a custom calibration, 0.1% accuracy is possible. The diamond shape (D-shape) tube design has the best overall performance of any Coriolis tube shape. The mechanical advantages of this design include the best signal-to-noise ratio and reduced effects of external vibrations, thus improving zero stability.
Mass Flow Rate (max.): | 650 kg/h | 24 lb/min |
Volumetric Flow Rate (max.): | 650 l/h | 2.86 gpm, 98 bbl/d (US) |
Standard Pressure Rating: | up to 200 bar | 2900 psi |
Nominal Meter Size: | DN4 | 1/8″ |
- Robust, no moving parts for long life
- Increased productivity, lower maintenance
- Custom flow connectors & installation lengths
- Lower installation cost
- Excellent repeatability (±0.05 % of flow rate)
- Increased product quality
- High Pressure Option: 345 bar (5,000 psi)¹
- Broad applications: dosing, paint, chemical injection
FLOW RATE | |
Mass Flow Rate (max.): | 650 kg/h, 24 lb/min |
Volumetric Flow Rate (max.): | 650 l/h, 2.86 gpm, 98 bbl/d (US) |
PRESSURE | |
Standard Pressure Rating: | up to 200 bar, 2900 psi (option: up to 345 bar, 5,000 psi)¹ |
ACCURACY (LIQUIDS)² | |
Basic Accuracy (Mass flow)³: | ±0.3 % (option: up to ±0.1 %) of flow rate |
Mass Flow Repeatability: | ±0.05 % of flow rate |
Basic Accuracy (Volume flow)³: | ±0.4 % (option: up to ±0.15 %) of flow rate |
Volume Flow Repeatability: | ±0.05 % of flow rate |
Zero Stability: | ±0.01 % of full scale |
ACCURACY (GASES) | |
Basic Accuracy (Mass flow)³: | ±1 % (option: up to ±0.5 %) of flow rate |
Repeatability (Mass Flow): | ±0.25 % of flow rate |
DENSITY | |
Density Range: | up to 2500 kg/m³, 2.5 g/cm³ |
Density Accuracy³: | ±1.0 kg/m³, ±0.001 kg/m³ |
Density Repeatability: | ±0.5 kg/m³, ±0.0005 g/cm³ |
TEMPERATURE | |
Process Temperature Range: | Standard: -40 °C … +100 °C -(40 °F … +212 °F ) Optional: -60 °C … +200 °C (-76 °F … +392 °F ) |
Temperature Accuracy: | ±1 °C ±0.5 % of reading (±1.8 °F ± 0.5% of reading ) |
Temperature Repeatability: | ±0.2 °C (±0.36 °F) |
MATERIALS OF CONSTRUCTION (wetted parts) | |
Tubes: | 1.4404/AISI 316L |
Case: | 1.4404/AISI 316L |
Flow splitter: | CF3M /AISI 316L |
Brazing Alloy: | BNi-2 |
EX CERTIFICATIONS | |
ATEX (Ex): | Zone 1: Group IIC or IIB, T2…T4 |
ATEX (Exn): | Zone 2: II 3G Ex nA IIC T2…T4 Gc |
IECEx (Ex): | Zone 1: Group IIC or IIB, T2…T4 |
cCSAus (Ex1): | Class I, Division 1: Group A, B, C, D or C, D, T2…T4 For Ex1, Ex3, EAC (TR-CU) the electronic is only available in aluminum housing. |
ATEX + IECEx + cCSAus triple approval (Ex3): | Zone 1: Group IIC or IIB, T2…T4 and Class I, Division 1: Group A, B, C, D or C, D, T2…T4 For Ex1, Ex3, EAC (TR-CU) the electronic is only available in aluminum housing. Only with remote electronics. |
EAC (TR-CU): | Group IIC or IIB, T2…T4 For Ex1, Ex3, EAC (TR-CU) the electronic is only available in aluminum housing. |
¹ The 200 bar standard pressure rating is per ASME. The 345 bar option is non-ASME rated.
² Stated flow accuracy combines the effects of repeatability, linearity and hysteresis.
³ The specifications refer to standard conditions (for further information see TCM CLASSIC Series Manual).
Accuracy For Liquids:
Standard (Single-Point Calibration, Basic Accuracy: ±0.3 %)
Custom (Five-Point Calibration, Basic Accuracy: ±0.1 %)
Accuracy For Gases:
Standard (Single-Point Calibration, Basic Accuracy: ±1 %)
Custom (Five-Point Calibration, Basic Accuracy: ±0.5 %)